| Model Number |
MMT-723RNAL |
| Medical Device Problem Code |
Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump revived with blank display due to moisture damage to the electronic assembly noted.Unable to perform displacement test, self test, off no power test, stop current test and run current test due to blank display.(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had expose to fluid.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|