| Model Number |
MMT-723NAH |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Pump passed the displacement test but motor error alarm during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to motor error alarm.Motor passed motor test.(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error alarm.Customer stated they were able to clear the alarm and they were able to rewind insulin pump.Customer stated drive support cap appears normal.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|