MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number ETLW1610C156E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Obstruction/Occlusion (2422)
Date of Event 07/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of an unknown size abdominal aortic aneurysm.Approximately 1 months post procedure, it was reported that a ct was performed, as the patient was complaining of left leg pain, where the ct showed what appeared to be a left limb occlusion.Thrombectomy was performed and a non-medtronic stents and a self expanding stenting were implanted in the limb 2 days later.As per the physician, cause of the event was patient's anatomy.The physician stated that the transition from the left common iliac to the left external was tortuous no additional clinical sequalae were reported and the patient is fine.
 
Additional Manufacturer Narrative
Product analysis the reported occlusion could not be confirmed on the limited films received; therefore the cause of the event could not be determined.Post-implant ct¿s were not available for review and the stent graft in-vivo configuration could not be evaluated.It is possible that patient anatomical factors may have been a factor in the reported occlusion but this could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key12196363
Report Number9612164-2021-02792
Device Sequence Number1377193
Product Code MIH
UDI-Device Identifier00643169467637
UDI-Public00643169467637
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/14/2022
Device Model NumberETLW1610C156E
Device Catalogue NumberETLW1610C156E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2021
Initial Date Received by Manufacturer 07/14/2021
Supplement Date Received by Manufacturer09/21/2021
Initial Report FDA Received Date07/20/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age72 YR
-
-