| Model Number |
C-MOUNT HIP AR-SCOPE 70°,3.5X 204MM |
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
06/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hip arthroscopic / labral reconstruction procedure, the distal tip of the scope broke off into the patient.No additional information provided.Additional information requested.Additional information: halfway through the case the surgeon discovered a piece of the scope on the x-ray machine inside the soft tissue.The surgeon made short attempt to retrieve it but was unsuccessful and continued with the procedure.At the very end of the case they spent 45 minutes to locate it inside the patient and additional x-rays were taken but they were unable to retrieve the fragment.The doctor used a non-arthrex shaver and burr which came in contact with the scope which severed part of the scope.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint was confirmed.Visual evaluation confirmed the condition of the distal tip fibers and welded metal tubing are broken.Broken piece was not returned.It was also observed, scratches around shaft at the distal end and around the lens of the device.Based on the device evaluation, the cause of the event is misuse by the end user, as stated in the complainant's narrative that the burr came in contact with the scope which severed part of the scope.
|
| |
|
Search Alerts/Recalls
|