Model Number 1103
Device Problem Increase in Suction (1604)
Patient Problem Arrhythmia (1721)
Event Date 07/11/2021
Event Type  Injury  
Manufacturer Narrative
Additional information has been requested regarding the details of the event, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.Investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced a cardiac arrhythmia and the ventricular assist device (vad) exhibited suction.The patient was shocked by their implantable cardioverter defibrillator (icd).An electrocardiogram was performed.The vad remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A supplemental report is being submitted for device investigation.Product event summary: the ventricular assist device (vad) was not returned for evaluation.The reported suction event could not be confirmed via review of the controller log files since log files were not available for analysis.Information received from the site revealed that the patient experienced a cardiac arrhythmia.The patient was shocked by implantable cardioverter defibrillator (icd) and an electrocardiogram was performed.Based on the available information, there is no evidence to indicate that a device malfunction or performance issue caused or contributed to the reported event.Based on the risk documentation, possible causes of the reported suction event may be attributed to multiple factors including but not limited to thrombus at the inflow cannula/outflow graft, constriction at the outflow graft, poor vad filling, and/or inappropriate pump rotational speed.Per the instructions for use, cardiac arrhythmia is a known potential complication associated with the implantation of an vad.There was no evidence that the patient had a history of similar adverse events.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A supplemental report is being submitted for additional information.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the patient was admitted and "tuned up" with standard assessment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th avenue
miami lakes FL 33014
MDR Report Key12196802
MDR Text Key262458968
Report Number3007042319-2021-05407
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707003261
UDI-Public00888707003261
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2021
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Date Manufacturer Received09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age62 YR
Patient Weight52
-
-