MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/35; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.5/35; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 391018
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Event or Problem Description
An orsiro drug eluting stent system was selected for treatment of a moderately calcified lesion (100 percent stenosis degree) in the mildly tortuous lad.It was tried to deliver the device but the orsiro did not cross the lesion.Once it was removed from the patients body and checked it, it was found that stent was curled up at the tip edge.
 
Additional Manufacturer Narrative
Combination product: yes.On 23-aug-2021 originally reported on wrong lot number.Closing this report and opened mdr-1028232-2021-04748 with the correct information.Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.5/35
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key12196887
Report Number1028232-2021-04066
Device Sequence Number19783822
Product Code NIQ
UDI-Device Identifier07640130439399
UDI-Public07640130439399
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/07/2022
Device Model Number391018
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03202763
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/16/2021
Supplement Date Received by Manufacturer08/23/2021
Initial Report FDA Received Date07/20/2021
Supplement Report FDA Received Date08/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-