| Model Number |
401756 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug eluting stent system was selected for treatment of a pre-dilated moderately calcified lesion (80 percent stenosis degree) in the moderately tortuous rca.The stent could not pass the lesion and was deformed at the distal end.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the complaint instrument nor the angiographic material was available for analysis.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the product release documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no manufacturing or material related root cause could be determined.
|
| |
|
Search Alerts/Recalls
|