Model Number 0998-00-0800-53
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed that the iabp unit would start power cycling when plugged into wall power or when the batteries were installed.The fse troubleshot and narrowed down the issue to the power management board.The fse replaced the power management board (p/n: 0670-00-1162).Additionally, the fse observed the reported printer issue and replaced the printer.Additionally, the touchscreen repeatedly failed to calibrate.These issues are unrelated to the reported issue and separate mdrs are being submitted for each.The fse ordered parts for the touchscreen and returned at a later date to complete repairs.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6).
 
Event Description
It was reported that during previously scheduled service performed by a getinge field service engineer (fse) the cardiosave intra-aortic balloon pump (abp) unit power cycles (turns on and off repeatedly) when power is supplied through batteries or wall power.There was no patient involvement, and no adverse event reported.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12196991
MDR Text Key262468494
Report Number2249723-2021-01570
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-