| Model Number |
Intermediate Surgical Splint |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The sales representative notified 3d systems after patient surgery that the intermediate surgical splint did not set the patient's occlusal anatomy into the correct intermediate occlusion.The surgeon was able to adjust accordingly to achieve the correct final occlusion during surgery; however, contacted 3ds with concerns of why there was such a large deviation present.3ds performed an investigation that digitally reviewed the virtual surgical planning and design files.It was determined the intermediate splints were designed off of stl files that did not represent the surgeon's requested intermediate occlusion for the patient.This information was provided to the surgeon.No adverse effects were reported after surgery.
|
| |
|
Event or Problem Description
|
|
The surgeon reported the vsp system intermediate splint did not match the planned intermediate position for the patient's maxilla for double jaw orthognathic surgery.
|
| |
|
Search Alerts/Recalls
|