• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMÉRIEUX, INC. VIRTUO, A UNIT; VIRTUO®, A UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMÉRIEUX, INC. VIRTUO, A UNIT; VIRTUO®, A UNIT Back to Search Results
Model Number 411660
Medical Device Problem Code Data Problem (3196)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in portugal notified biomérieux of experiencing an issue with their bact/alert® virtuo¿ a unit (ref 411660, serial (b)(4)) in which the final negative result for four (4) bottles belonging to four (4) different patients did not transfer to myla® and the customer's lis (laboratory information system).The bottles were flagged negative on (b)(6) 2021 at 01:03 am and the virtuo automatically unloaded the bottles at 01:04 am.The final negative results were not sent to myla or the customer's lis.Local customer service (lcs) confirmed that there was a delay of 36 hours in reporting the final negative results for these bottles.Local customer service sent the negative results from the virtuo instrument and the results updated in myla and the customer's lis from negative to date (ntd) to final negative.Lcs confirmed that this event did not lead to any adverse event related to any patient's state of health.A biomérieux internal investigation will be initiated.
 
Additional Manufacturer Narrative
An investigation was conducted for one customer complaint for bact/alert® virtuo® serial number (b)(6) version r3 in a site configuration with myla® server version 4.7.1.After the monthly reboot of the virtuo pc at the first of the month on (b)(6) 2021, the customer noticed four bottle records did not update in myla to final negative, see list of the bottle ids below.Prwcwwzt, arzdrg84, arzdg8d7, nryh4bpv.The investigation could not identify the root cause or reproduce the issue with in-house testing.The most likely root cause would be an issue with the myla connector component related to a race condition with the myla monthly reboot and virtuo result messages occurring at the same time.The issue is isolated, and low risk to the patients as any records in myla missing their final result can be updated by pressing the upload button at virtuo.In this complaint case the bottle results were negative.If the bottle results were positive, the positive bottles would be unloaded from virtuo and the gram stain resulted in the lis and possibly called to the physician; therefore also a low risk to impact the patient.The customer noticed the four bottle records still showed a result of negative to date in myla, when they should have been final negative.Their immediate action was to call local customer service, who advised them to resend the results by pressing the upload button on the virtuo bottle record.No specific harm or impact to the two patients was stated.The bact/alert virtuo user manual was reviewed by the investigator and found to have adequate directions for the user.The manual describes use of the upload button to resend bottle records to myla.Queries of the manufacturing data and the complaint data do not reveal any adverse trend.For the complaint data the error code virtuo/myla interface problem - k056 was reviewed.This error code may be used for various issues, but no other complaints for records not updating after the first of the month virtuo pc reboot were found in the last 13 months.Global customer service recommends the customer check the myla results daily and press the upload button on virtuo¿s bottle record when they notice any results not updated on myla.Pending negative to date bottle results past their maximum test time can be targeted.Global customer service also recommends the customer notify biomérieux if the issue reoccurs.Biomérieux will continue to monitor complaints for this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIRTUO, A UNIT
Common Device Name
VIRTUO®, A UNIT
Manufacturer (Section D)
BIOMÉRIEUX, INC.
595 anglum road
hazelwood MO 63042
MDR Report Key12197490
Report Number1950204-2021-00045
Device Sequence Number3036738
Product Code MDB
UDI-Device Identifier03573026369767
UDI-Public03573026369767
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K161816
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number411660
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2021
Supplement Date Received by Manufacturer09/15/2021
Initial Report FDA Received Date07/20/2021
Supplement Report FDA Received Date09/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-