• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SERVICE REPL CAMERA HEAD LENS INT SYS; ARTHROSCOPE Back to Search Results
Model Number 72203967S
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that during use, outside the patient, the camera head lens int sys has black screen.The procedure was finished with a smith and nephew backup device.No surgical delay.Patient injuries were not reported.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was returned to the designated complaint unit for independent evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device did not identify any issues.Functional evaluation revealed the device was not recognized when plugged in.Further investigation revealed an internal damage to the cord.The complaint has been confirmed and the root cause has been associated with an electrical component failure.Factors, unrelated to the manufacturing and design of the device that could have contributed to the reported event include a damaged cord.A review of the device history records showed there were no indications to suggest that the device did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE REPL CAMERA HEAD LENS INT SYS
Common Device Name
ARTHROSCOPE
MDR Report Key12198041
Report Number1643264-2021-02215
Device Sequence Number17355549
Product Code HRX
UDI-Device Identifier00885556631980
UDI-Public00885556631980
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K153606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72203967S
Device Catalogue Number72203967S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Initial Date Received by Manufacturer 07/01/2021
Supplement Date Received by Manufacturer10/26/2021
Initial Report FDA Received Date07/20/2021
Supplement Report FDA Received Date10/27/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
-
-