• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES PRODUKTIONS GMBH SHAFT FOR TREPHINE ATTACHMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES PRODUKTIONS GMBH SHAFT FOR TREPHINE ATTACHMENTS Back to Search Results
Catalog Number 03.111.030
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2021 that the tip of the instrument broke off during surgery.There was no patient harm or surgical delay.The surgery was completed successfully with another instrument.This report is for one (1) shaft for trephine attachments.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: part: 03.111.030 lot: 47p9558 manufacturing site: bettlach.Release to warehouse date: 23 june 2020.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Visual inspection: shaft f/trephine+extraction attachment (part# 03.111.030, lot# 47p9558, qty# 1) was returned and received at us customer quality (cq).Upon visual inspection at cq, it is observed that the tip of the device was broken, and the broken piece was not returned no other issues were identified also pictures were reviewed.Device failure / defect identified? yes.Dimensional inspection: dimensional inspection of the received device was not performed at cq due to post manufacturing damage.Document/specification review: the following document(s) was reviewed: anschluss connector trepane orthopedic foot.No design issues or discrepancies were found during this investigation.Complaint confirmed? yes.Investigation conclusion: the complaint is being confirmed for shaft f/trephine+extraction attachment (part# 03.111.030, lot# 47p9558, qty# 1).While a definitive root cause could not be identified for the reported problem, it is possible that the device might have encountered unintended forces.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based upon these results, no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHAFT FOR TREPHINE ATTACHMENTS
Common Device Name
TREPHINE
Manufacturer (Section D)
SYNTHES PRODUKTIONS GMBH
muracherstrasse 3
bettlach 2544
SZ  2544
MDR Report Key12199188
Report Number9612488-2021-00006
Device Sequence Number14625152
Product Code HWK
UDI-Device Identifier07611819375901
UDI-Public07611819375901
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.111.030
Device Lot Number47P9558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Initial Date Received by Manufacturer 07/07/2021
Supplement Date Received by Manufacturer07/07/2021
Initial Report FDA Received Date07/20/2021
Supplement Report FDA Received Date09/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-