Model Number 3542615
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/01/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis deflation.Explantation month, day, and year: (b)(6) 2021.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female patient underwent a breast augmentation revision procedure with a mentor siltex round moderate profile 150cc saline breast prosthesis that deflated after implantation.Deflation of the patient¿s right breast prosthesis was diagnosed during a physical consultation.As a result, the patient is scheduled to undergo explantation on (b)(6) 2021.
 
Additional Manufacturer Narrative
On (b)(6) 2021, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 19-oct-2021, mentor received additional information about the event.The patient did not replace her explanted breast prostheses.On 04-nov-2021, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced deflation in the breast implant.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant had an area of siltex cracking on the posterior view.In addition, a tear was noted within the siltex cracking, measuring approximately 0.5 cm.The evaluation determined that the cause of the deflation is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.A second product was received (lot #6663283).No adverse events were reported for this concomitant (contralateral) device.Therefore, no further analysis is required.It should be noted that as part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key12199813
Report Number1645337-2021-08161
Device Sequence Number16514213
Product Code FWM
UDI-Device Identifier00081317001720
UDI-Public00081317001720
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2016
Device Model Number3542615
Device Catalogue Number3542615
Device Lot Number6663283
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2021
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 06/30/2021
Supplement Date Received by Manufacturer10/07/2021
10/19/2021
Initial Report FDA Received Date07/20/2021
Supplement Report FDA Received Date10/13/2021
11/07/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-