| Model Number |
M00545150 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During procedure, the cutting wire of the autotome rx 49 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 49 was analyzed, and a visual evaluation noted that the cutting wire was broken, blackened, and the distal part of the cutting wire was kinked which were consistent to findings when the device was observed under magnification.A functional evaluation was not performed due to the condition of the cutting wire.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, blackened, and the distal part of the cutting wire was kinked.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been caused due to energizing the device prior to performing sphincterotomy.This can compromise the cutting wire's integrity and lead to premature cutting wire fatigue.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 49 was used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2021.During procedure, the cutting wire of the autotome rx 49 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|