Manufacturing review: a device history record review could not be performed as the lot number is unknown.Investigation summary:the device was not returned for evaluation.Medical records were provided and reviewed.There were no device deficiencies identified within the medical records.Therefore, the investigation is inconclusive as no objective evidence has been provided to confirm any alleged deficiency with the filter.Additionally, it can be confirmed that the patient experienced pe post deployment.However, the relationship to the filter is unknown.Based on the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.
|