• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Component Missing (2306); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/18/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint rt380 adult dual-heated evaqua2 breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6) that a rt380 adult dual-heated evaqua2 breathing circuit did not have heater wires inside the expiratory limb.There was no patient involvement.
 
Event Description
A distributor reported on behalf of a healthcare facility in south korea that a rt380 adult dual-heated evaqua2 breathing circuit did not have heater wires inside the expiratory limb.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint rt380 adult evaqua2 breathing circuit was not returned to fisher & paykel healthcare (f&p) in new zealand for investigation.Our investigation is thus based on the information and photograph provided by the customer, and our knowledge of the product.Results: visual inspection of the provided photograph revealed that the expiratory limp did not have a heater wire and heater wire plug.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.All rt380 breathing circuits are visually inspected and pressure tested for leaks before releasing for distribution.Any breathing circuit which fails any of these tests is discarded.If any faults are detected the whole batch is placed on hold for investigation.The subject breathing circuit would have met the requirements at the time of production.The user instructions that accompany the rt380 breathing circuit state the following: - check all connections are tight before use.- set appropriate ventilator alarms.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers.- do not reuse this product.Reuse may result in transmission of infectious substances, interruption to treatment, serious harm or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
MDR Report Key12201468
MDR Text Key264945510
Report Number9611451-2021-00852
Device Sequence Number1
Product Code BZE
UDI-Device Identifier09420012429728
UDI-Public(01)09420012429728(1002101529841(11)210227
Combination Product (y/n)N
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot Number2101529841
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received07/21/2021
Supplement Dates Manufacturer Received09/09/2021
Supplement Dates FDA Received10/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-