| Model Number |
CSE-P-125-JA |
| Medical Device Problem Codes |
Crack (1135); Mechanical Problem (1384); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2021, haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set 125ml from (b)(6) and confirmed blood in the inner core with a crack in the inner core base.Visual inspection confirmed.The blood was seen inside the core in the bowl.Bowl decompression test was performed:.The blood outflow from the inner core and the outer core welding part was confirmed.Cause of the event is due to manufacturing.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, haemonetics was notified of a long hour returning error which was observed during a procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 125ml.There was no reported impact to patients' health.
|
| |
|