• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ENCOR BIOPSY PROBE; BIOPSY INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ENCOR BIOPSY PROBE; BIOPSY INSTRUMENT Back to Search Results
Model Number ECP0112G
Medical Device Problem Codes Filling Problem (1233); Suction Problem (2170); Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 09/2021).
 
Event or Problem Description
It was reported that during a preparation for a breast biopsy, the device allegedly had a suction issue.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one biopsy probe were returned for evaluation.During visual evaluation, the probe appeared clean and no other anomalies were noted on the device.It was noted that foreign material appeared on the trocar tip.An enspire system and in-house saline cap were used for functional testing.Upon functional testing, the probe was loaded into the driver and calibrated successfully.The probe passed in maximum vacuum test and failed in vacuum differential test.The probe was inserted into a piece of beef and around the clock sampling was performed.The probe was able to obtain 6 samples successfully.Therefore, the investigation is confirmed for the reported foreign material and identified filling issue as foreign material was found on the trocar tip and the probe fails to meet the requirement of vacuum differential test.However, the investigation is unconfirmed for the reported suction issue as the device meets the requirement of maximum vacuum test.A definitive root cause for the reported failure could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a preparation for a breast biopsy, the device allegedly had a suction issue.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR BIOPSY PROBE
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12202292
Report Number2020394-2021-01361
Device Sequence Number5061831
Product Code KNW
UDI-Device Identifier00801741086267
UDI-Public(01)00801741086267
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K051158
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberECP0112G
Device Catalogue NumberECP0112G
Device Lot NumberVTDW0659
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer07/27/2021
Initial Report FDA Received Date07/21/2021
Supplement Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-