BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problems Entrapment of Device (1212); Mechanical Problem (1384); Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2021
Event Type  malfunction  
Event Description
It was reported that the blades stopped spinning, and the device became stuck on the guidewire.A 2.4mm jetstream xc atherectomy catheter was selected for use in an atherectomy procedure in the popliteal and superficial femoral artery to treat peripheral artery disease.There was some tortuous anatomy in the target location.During the procedure in the middle of a run through a chronic total occlusion, significant resistance was noted followed by loss of rotation while in blades up mode.Attempts were made to turn the device off and on as well as advance and retract, but the jetstream device was bound to the non-boston scientific guidewire.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12203083
MDR Text Key262666028
Report Number2134265-2021-09320
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889922
UDI-Public08714729889922
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/13/2023
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0027121152
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MEDTRONIC 014 6 SPIDER FILTER WIRE
-
-