Model Number 1020-35-116
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional product code: kwp complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2021, during posterior spinal fusion surgery there were multiple issues while using symphony.It was noticed that the screw head chip when a set screw was apparently cross threaded while using the reduction gun.It was unknown how it occurred.Also, there were two set screw drivers that had tips sheared off while inserting set screws.Fragments were generated and easily removed.There was a surgical delay of ten (10) minutes.There were no patient consequences.The procedure was successfully completed.Concomitant device reported: unknown reduction gun (part# unknown, lot# unknown, quantity unknown) this complaint involves four (4) devices.This report is for (1) 4.0 ply scrw 3.5x16 this report is 1 of 4 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.0 PLY SCRW 3.5X16
Common Device Name
ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key12203191
Report Number1526439-2021-01441
Device Sequence Number5061224
Product Code NKG
UDI-Device Identifier10705034508702
UDI-Public10705034508702
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K181949
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1020-35-116
Device Catalogue Number102035116
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2021
Initial Report FDA Received Date07/21/2021
Patient Sequence Number1
-
-