Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.The instrument was found to have a broken conductor wire at the weld joint of the yaw pulley after one side of the clevis ear was removed for further investigation.The silicone potting appear to be compromised and the conductor wire thermally damaged around the weld location.Any material missing is likely to be thermally induced rather than mechanically induced.Failure analysis found the primary failure of broken conductor wire at the weld to be related to the customer reported complaint.Root cause is attributed to a device design.Additional observation not reported by the site was that the instrument was found have thermal damage of the monopolar yaw pulley.Black char marks were observed.Any material missing is likely to be thermally induced rather than mechanically induced.Failure analysis found the additional failure of thermal damage of the monopolar yaw pulley-weld damage to be related to the customer reported complaint.Root cause is attributed to a device design.A site history complaint review was performed and no related complaints were found to the product.A system log review confirmed the permanent cautery hook was last used on (b)(6) 2021 on (b)(4) with 4 uses remaining.The permanent cautery hook is intended to be used with the da vinci system for precise dissection and division of tissue with monopolar cautery.The instrument is designed to provide energy from the designated location on the instrument (the tip) to the planned anatomical location when used as intended.The energy is activated by pressing the designated pedal on the surgeon side console (ssc).This complaint is reportable due to the following conclusion: there was evidence of thermal damage to the weld with no indication or claim of user mishandling or misuse.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
|