MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9736242
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2021
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A software analysis was initiated.Analysis found that the behavior was consistent with a known software anomaly in the medtronic navigation software anomaly tracking database.Concomitant medical products: product id: 9736226, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the navigation system displayed an error 64 message.Prior to the error message, it was noted that the navigation system was running slowly following a first patient registration.Following an additional registration, the error message appeared.It was noted that subsequent procedures for the day were completed without navigation from the outright.There was no patient present when this issue was identified.The reported issue was suspected to have been caused by corrupt patient exams.Additional information was received stating that there was a patient present when the reported issue occurred.It was unknown what type of procedure was being performed.The surgeon completed the procedure with the use of the navigation system.
 
Event Description
Additional information was received stating that the manufacturer representative talked with the site and stated to follow the scan protocol but that they also need to keep an eye on this issue.Because this could be an issue with the software and it was recommended to upgrade the software.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: product id: 9736226, version #: 1.3.2 h3) the software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
The manufacturer representative went to the site to test the navigation system.No hardware parts were replaced.The software was upgraded.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION FLEXENT
Type of Device
EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key12205860
MDR Text Key262870676
Report Number1723170-2021-01886
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
PMA/PMN Number
K200723
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9736242
Device Catalogue Number9736242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Date Manufacturer Received10/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."
-
-