MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 CHANNEL DRAIN, 10FR., HUBLESS FULL FLUTE WITH TROCAR; SILICONE CHANNEL DRAIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 CHANNEL DRAIN, 10FR., HUBLESS FULL FLUTE WITH TROCAR; SILICONE CHANNEL DRAIN Back to Search Results
Model Number 072227
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bard drain was broken from the trocar inside the patient's breast pocket.
 
Additional Manufacturer Narrative
The reported event was confirmed however the cause was unknown.One sample was confirmed to exhibit the reported failure.The device did not meet the specifications.The product was used for the patient treatment.The product had caused the reported failure.Visual evaluation of the returned sample noted one opened (without original packaging) used channel drain received with trocar.Visual inspection of the sample the channel drain was broken off from the trocar.There was a piece of the drain still attached to the trocar given evidence of the break.A potential root cause for this failure mode could be due to the drain was manufactured incorrectly (dimensions or material out specifications, curing time not appropriate, etc.).The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "indications: wound drains are used to remove exudates from wound sites.Warnings: an effective closed suction drain system requires maintenance of the system to preserve patency.The drain must not be allowed to occlude nor the reservoir to completely fill; and reservoir suction must be maintained in order for the system to function properly.Verify that the system is functioning properly.If the system is not maintained properly, surgical complications, including hematomas, may result.In the event of occlusion of the drain, all wound drainage via the drain ceases.Careful attention to the drain will minimize the possibility of this problem.If occlusion does occur, the drain can be aspirated by connecting suction to the reservoir outlet or temporarily disconnecting the drain from the reservoir and applying suction directly to the drain.If an air-tight seal between the drain and the skin where the drain emerges is not achieved, the air leak must be rectified or the system must be converted to open drainage.An airtight seal between all system components (drain, adaptor and reservoir) is necessary for proper system function.Leaving the soft silicone elastomer drain implanted for any period of time so as to cause tissue ingrowth around the drain can interfere with easy removal and may effect the performance of the drain.The surgeon should monitor the patient¿s rate of wound healing.Evacuators should be used in cardio-thoracic surgery only after the lung is fully expanded and all air leaks have sealed.Drain perforations or channels must lie within the wound or cavity to be drained, otherwise inadequate drainage may result.To avoid the possibility of drain damage or breakage: avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.Drains should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Complications: complications which may result from the use of this suction drainage system include the risks associated with methods utilized in the surgical procedure, as well as the patients degree of intolerance to any foreign object in the body.The advantages of wound drainage, particularly closed system drainage, are lost if an air-tight seal between the drain and the skin where the drain emerges is not achieved, or if the drain is allowed to become occluded or if the reservoir is not activated properly, doesn¿t function properly or is not monitored.Evacuators should be emptied and re-activated when required per hospital protocol.From the leak; subsequent drainage to the reservoir will occur only if allowed by gravity and wound exudate forcing the flow.Entry into the reservoir is allowed only by displacement of air in the reservoir by wound exudate flow.In this displacement process, air reflux from the reservoir to the wound can occur and increase the likelihood of back-contamination across the anti-reflux valve.In the event of drain occlusion by fibrin, clots, or other particulate matter, all wound drainage via the drain ceases.If the reservoir is not emptied when it is full, equilibrium between the drain and reservoir at wound pressure will ultimately occur and drainage from the wound site will cease.When the reservoir and drain are at the same pressure and the reservoir is full of fluid, the likelihood of back-contamination across the anti-reflux valve is increased.When used to drain the pleural cavity in the presence of an air leak, drains must be attached to an appropriate pleural cavity drainage system to prevent tension pneumothorax.Drain placement: the surgeon should irrigate the wound with sterile fluid, then suction the irrigating fluid and gross debris from the operative site.Tubes should lie flat and in line with the anticipated skin exit.To facilitate later removal by manual traction, the tubing should not be curled, pinched, or sutured internally.Positioning of the drain in the body cavity, as well as the number of drains indicated, should be determined by the operating surgeon.Drain tubing should be placed within the wound by approximating the areas of critical fluid collection.Care must be taken to ensure that all drain perforations or channels lie completely within the wound or cavity to be drained.Taping or a triple loop suture (around and not through the tubing) will aid in preventing accidental drain displacement.Deep drainage is best accomplished by using one or more drains for each level of tissue.Each level should be evacuated by a separate vacuum source.Care must be exercised to avoid damage to the drain (see warnings).The tubing should be repeatedly checked during closure for free motion to avoid breakage and/or fragment retention within the wound.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that the bard drain was broken from the trocar inside the patients breast pocket.Per additional information received via email on 30aug2021 it was stated that the drain was disconnected from the trocar during insertion and was on a scalp.Per notification from investigator on 18jan2022 the sample evaluation confirmed that the channel tube broken from the trocar and stated there was only breakage and no disconnection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHANNEL DRAIN, 10FR., HUBLESS FULL FLUTE WITH TROCAR
Common Device Name
SILICONE CHANNEL DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key12206807
Report Number1018233-2021-04406
Device Sequence Number4486470
Product Code GBX
UDI-Device Identifier00801741049873
UDI-Public(01)00801741049873
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model Number072227
Device Catalogue Number072227
Device Lot NumberNGEW1003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/28/2021
Supplement Date Received by Manufacturer01/24/2022
Initial Report FDA Received Date07/21/2021
Supplement Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-