| Model Number |
1210-6450 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
07/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that during a gamma nail procedure, a k-wire broke of in the patient while performing femoral fixation.Surgeon decided to leave the broken portion in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that during a gamma nail procedure, a k-wire broke of in the patient while performing femoral fixation.Surgeon decided to leave the broken portion in the patient.
|
| |
|