Model Number 1012580-26
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/15/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis as the stent remains in the patient.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported failure to seal the perforation; however, the subsequent treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.The additional device (other graftmaster) referenced is being filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that on (b)(6) 2021 the patient underwent a percutaneous coronary intervention in the distal right coronary artery (rca).A small perforation occurred during the procedure.The 2.8x26 mm graftmaster stent was implanted, but the leak continued.The second 2.8x26 mm graftmaster stent was implanted, but the leak persisted.Coils were implanted and successfully sealed the perforation.There were no issues with the graftmaster besides the perforation not being sealed.There was no adverse patient sequela.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER RX CORONARY STENT GRAFT SYSTEM
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12206949
Report Number2024168-2021-06329
Device Sequence Number17041708
Product Code MAF
UDI-Device Identifier08717648176364
UDI-Public08717648176364
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number1012580-26
Device Catalogue Number1012580-26
Device Lot Number0052241
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/06/2021
Initial Report FDA Received Date07/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight88
-
-