| Model Number |
PRD-SC30-125 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
06/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
| |
|
Event or Problem Description
|
|
A stealth peripheral orbital atherectomy device (oad) was used to treat a chronic total occlusion lesion in the superficial femoral artery via femoral access.Two treatments were administered on low speed.The oad became stuck on the guide wire during removal.The wire and oad were removed together, and wire access was lost.The procedure was delayed by 30 minutes due to difficulty rewiring the vessel.A flow-limiting dissection was observed in the area where the oad had become stuck on the guide wire.An unplanned stent was deployed over the dissection.
|
| |
|