| Model Number |
TPS005 |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample was returned for evaluation.A follow-up report will be submitted once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
Two foreign bodies were noticed on post stent deployment imaging.A foreign body was noticed in the heart and jailed between the stent and body.The tip of the 10f sheath did not look intact on imaging.When the sheath was removed, a the distal end was not whole and a section fragmented.
|
| |
|
Event or Problem Description
|
|
Two foreign bodies were noticed on post stent deployment imaging.A foreign body was noticed in the heart and jailed between the stent and body.The tip of the 10f sheath did not look intact on imaging.When the sheath was removed, a the distal end was not whole and a section fragmented.
|
| |
|
Additional Manufacturer Narrative
|
|
One opened 10f introducer sheath was returned for evaluation.Visual inspection of the sheath established that the tip was cracked, and a piece was missing, thereby confirming the complaint.The sheath¿s original blunt end design was changed to a new dual-taper tip design to improve the ease of passing the tip through the parenchyma.The od and the id of the tube were increased by 0.003¿, and the tip geometry was changed from a single radius blunt tip to a dual-taper sharp tip that resulted in a thinner wall thickness at the most distal tip of the catheter.The potential root cause for the sheath tip failure has not yet been conclusively determined.It is theorized that it might be the result of the material used, processing parameters of the material, and/or the new tip geometry.Capa c-2021-046 has been initiated to capture the root cause investigation.No product currently exists in the field while this investigation is being conducted.
|
| |
|