| Model Number |
45-480815-L |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for evaluation.The alleged product issue could not be confirmed.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during embolization of a splenic artery embolization, the physician encountered resistance during advancement.Upon retraction, the coil had detached.The coil was removed without intervention.There was no reported patient injury.
|
| |
|