Catalog Number 955608
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/30/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that five minutes after starting treatment with a polyflux 140h, an external blood leak at the arterial header was observed.There was patient involvement however, no patient injury or medical intervention was reported.No additional information is available.
 
Manufacturer Narrative
Additional information added to d9, h3, h6, h10 h10; the actual device was received for evaluation.The visual inspection by naked eye of the product did not show any conspicuousness.A leak test of the blood side and the dialysate side was performed and no leak was observed.A leak test with bovine blood was performed and no leak was observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 140H
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12210453
MDR Text Key262942171
Report Number9611369-2021-00174
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124526
UDI-Public(01)07332414124526
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955608
Device Lot Number0-5502-H-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Date Manufacturer Received08/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
TORAY TR-3000MA; TORAY TR-3000MA
-
-