U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEX03 MEXICO-JUAREZ PRESOURCE OPEN HEART PACK DRP SH 36883 (ODSNC)654; CARDIOVASCULAR PROCEDURE KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEX03 MEXICO-JUAREZ PRESOURCE OPEN HEART PACK DRP SH 36883 (ODSNC)654; CARDIOVASCULAR PROCEDURE KIT Back to Search Results
Model Number SCVHDODSNC
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint was forwarded to the manufacturing facility where it is currently still under investigation.A follow-up report will be filed once the results have been completed.
 
Event or Problem Description
Customer reported that during a coronary artery bypass graft procedure, a piece of lint was found on the towel pwtb04-stm from the open heart pack scvhdodsnc.Towel used for holding instruments.There was no injury to the patient.
 
Additional Manufacturer Narrative
Supplemental report is being filed since the results of investigation are available.From the device history record, lot# 20210120-23-sh was finished on 26th jan 2021.No exception was recorded in the device history record that could lead to the reported incident.Based on supplier investigation, device history record (dhr) review did not indicate any exception that could lead to the reported incident.Supplier checked on their retained sample, the sample is intact without any lint on it.No physical sample was available for investigation.A photo was provided showing blue lint on the towel.According to the supplier, before production operation, the staff shall clean the equipment according to requirements and clean the lint on the equipment before operation.Supplier performed suction after the sewing, which can remove the foreign matters on the towel.The workers also do the visual test for every piece of product when folding.The lint in the complaint photo can be removed by employees, and it can also be removed in the suction process.In the folding process, supplier used one cloth pad under 100 pcs semi-finished products to avoid linting stuck onto the products during products' transfer.In the production process, employees clean the worktable every 2 hours.The foreign matter is always being controlled as a main defect factor.The process sampling criteria is aql:0.25.For the final product, it is i, aql:0.65.They checked the dhr, no record of foreign matter was found.Supplier confirmed this is an isolated incident and the root cause cannot be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN HEART PACK DRP SH 36883 (ODSNC)654
Common Device Name
CARDIOVASCULAR PROCEDURE KIT
Manufacturer (Section D)
MEX03 MEXICO-JUAREZ PRESOURCE
c. arcadia 1580 , col. terraza
cd.juarez , chihuahua CP 32 599
MX  CP 32599
MDR Report Key12211253
Report Number1423537-2021-00655
Device Sequence Number20108167
Product Code OEZ
UDI-Device Identifier10887488787343
UDI-Public10887488787343
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCVHDODSNC
Device Catalogue NumberSCVHDODSNC
Device Lot Number629280
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/28/2021
Supplement Date Received by Manufacturer06/28/2021
Initial Report FDA Received Date07/22/2021
Supplement Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-