• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL DUAL LF 6; BOTTLE, COLLECTION, VACUUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PE SAHARA DRY SUCT/DRY SEAL DUAL LF 6; BOTTLE, COLLECTION, VACUUM Back to Search Results
Model Number IPN028485
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Air leak in the red-bleu connectors with clamp above connection there is an air leak with clamp below connection there is no air leak so it is in the connection.
 
Event or Problem Description
Air leak in the red-bleu connectors with clamp above connection there is an air leak with clamp below connection there is no air leak so it is in the connection.
 
Additional Manufacturer Narrative
(b)(4).A visual inspection of the product involved in the complaint could not be conducted since the product was not returned.A functional inspection of the product involved in the complaint could not be conducted since the product was not returned.No additional tests were performed as part of this complaint investigation.The device history record of batch number 74f2000978 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled and inspected according to our specifications.No corrective action can be established at this moment since the product sample is not available for evaluation.Product sample is necessary to perform a proper investigation to determinate the root cause and the corresponding corrective actions.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility neither is being manufactured at the time.If defective sample becomes available at a later date this complaint will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PE SAHARA DRY SUCT/DRY SEAL DUAL LF 6
Common Device Name
BOTTLE, COLLECTION, VACUUM
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key12213345
Report Number3004365956-2021-00247
Device Sequence Number11565674
Product Code KDQ
UDI-Device Identifier14026704631671
UDI-Public14026704631671
Combination Product (Y/N)N
Initial Reporter CountryNL
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberIPN028485
Device Catalogue NumberS-1102-08LF
Device Lot Number74F2000978
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/21/2021
Supplement Date Received by Manufacturer07/28/2021
Initial Report FDA Received Date07/22/2021
Supplement Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-