(b)(4).Concomitant devices -vngd ps open por fmrl lt 62.5, catalog #: 184526, lot #: 993610; e1 vngd ps+ tib brg 63/67x10, catalog #: ep-183720, lot #: 315430; biomet finned pri stem 80x10mm, catalog #: 141316, lot #: 486030; bmet regenx pri tib tray 67mm, catalog #: 141272, lot #: 060430.The complainant has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The results of the investigation are as follows: no devices or photos were received; therefore, the condition of the components is unknown.Review of the device history record identified no related deviations or anomalies during manufacturing.Primary operative notes does not suggest any intra operative complications.Office visit notes from the doctor states that x-ray report revealed that all three patella pegs are broken causing loose patella and pain.The root cause of the patella peg fracture was determined to be a labeling and training deficiency as the surgical technique addendum for installation of regenerex patella failed to include guidance on depth line indication on the peg drill bit.A corrective action has been initiated to address the malfunction.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|