• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DII CONTROLLER; ARTHROSCOPE Back to Search Results
Catalog Number 72200873
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that the dii controller was running hot.The incident occurred before the procedure.There was no delay.It is unknown if there was an available back up device.No other complications were reported.
 
Manufacturer Narrative
Internal complaint reference (b)(4).The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with alarm sounding and "short circuit detected" error message on port a only.Cause of alarm and error is shorted electronic components on the main digital pcb.Resistor r54 was observed to be burnt.Product passed functional testing with a known good main pcb installed.The complaint investigation has concluded the cause of the failure to be defective electronic components.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DII CONTROLLER
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key12214342
MDR Text Key263140128
Report Number1643264-2021-02224
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K062849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2021
Date Manufacturer Received08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-