| Model Number |
7134 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an emerge mr balloon catheter.The device was microscopically and visually examined.At 81cm from the strain relief the hypotube was kinked.At 85.2cm from the strain relief there was separation of the hypotube.The distal portion of the separation there was a shaft kink at 46.6cm.There were kinks in the shaft on both ends of the proximal markerband in the balloon.There was proximal markerband damage.There was contrast present in the inflation lumen and the balloon was tightly folded.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 02-jul-2021.It was reported that shaft kink occurred.The 80% stenosed target lesion was located in the proximal part of left anterior descending artery.A 2.50mm x 20mm emerge balloon catheter was selected for use.However, during preparation, it was found out that the connecting part of the balloon was kinked.The procedure was completed with another of same device.No patient complications were reported and the patient status is stable.However, device analysis revealed markerband damage and hypotube break.
|
| |
|