• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1804275-12
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material rupture; however, factors that may contribute to material ruptures include, but are not limited to, material damage, interactions with other devices or interaction with lesion calcification and tortuosity.The product risk assessment identifies this as a foreseeable event; as such, design and manufacturing controls and mitigation's have been established for potential causes.There was no damage or leak noted to the stent delivery system (sds) during the inspection prior to use or during preparation of the device, which suggests a product quality issue did not contribute to the reported difficulties.Based on the information provided, a definitive cause for the reported material rupture cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the patient presented with an acute myocardial infarction (ami) and the procedure was to treat the left anterior descending (lad) coronary artery with moderate calcification, moderate tortuosity and 90% stenosis.Intravascular ultrasound (ivus) was performed and the 2.75x12 mm xience skypoint stent was placed.The balloon was inflated once at 5 to 6 atmospheres (atm) for 5 seconds and ruptured.The stent delivery system was removed without resistance and the stent was confirmed with angiography to be deployed.Post-dilatation was performed about 3 times with a 2.75x10mm balloon and it was confirmed with ivus that the stent was completely apposed and the procedure was completed.It was reported that the stent easily crossed and was apposed as it was short length stent.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12214891
Report Number2024168-2021-06379
Device Sequence Number3036009
Product Code NIQ
UDI-Device Identifier08717648233142
UDI-Public08717648233142
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2023
Device Model Number1804275-12
Device Catalogue Number1804275-12
Device Lot Number0090241
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2021
Initial Report FDA Received Date07/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE CATHETER: HEARTRAIL 6F; GUIDE WIRE: SION BLUE
-
-