| Model Number |
80618 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the outer shaft ruptured.A direxion microcatheter was selected for use in an embolization procedure.During the procedure, a loop formed, and the outer shaft ruptured in the middle of the catheter, outside the patient.The device was removed from the patient intact.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device eval by manufacturer: the direxion microcatheter was returned to boston scientific for analysis.The device shaft was analyzed for any damage.The shaft was fractured in two locations: 17.7cm and 62cm from the hub.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Fractures were confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that the outer shaft ruptured.A direxion microcatheter was selected for use in an embolization procedure.During the procedure, a loop formed, and the outer shaft ruptured in the middle of the catheter, outside the patient.The device was removed from the patient intact.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|