• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DIREXION; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DIREXION; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 80618
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the outer shaft ruptured.A direxion microcatheter was selected for use in an embolization procedure.During the procedure, a loop formed, and the outer shaft ruptured in the middle of the catheter, outside the patient.The device was removed from the patient intact.The procedure was completed with another of the same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device eval by manufacturer: the direxion microcatheter was returned to boston scientific for analysis.The device shaft was analyzed for any damage.The shaft was fractured in two locations: 17.7cm and 62cm from the hub.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Fractures were confirmed.
 
Event or Problem Description
It was reported that the outer shaft ruptured.A direxion microcatheter was selected for use in an embolization procedure.During the procedure, a loop formed, and the outer shaft ruptured in the middle of the catheter, outside the patient.The device was removed from the patient intact.The procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12215048
Report Number2134265-2021-09394
Device Sequence Number19908839
Product Code KRA
UDI-Device Identifier08714729839354
UDI-Public08714729839354
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2023
Device Model Number80618
Device Catalogue Number80618
Device Lot Number0027340086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/12/2021
Supplement Date Received by Manufacturer11/09/2021
Initial Report FDA Received Date07/22/2021
Supplement Report FDA Received Date11/22/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-