Model Number 10-85-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue during extensive testing of the device.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova (b)(4) received a report that a s5 double head pump used as sucker pump gave an error message.Reportedly it happened several times during procedure.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H.10: the pump serial read out analysis revealed that on the date of the event three errors occurred.Possible root causes are: occlusion set too hard by the user.Wrong tubing size.Rotor blocked by foreign object.Bearing defective preventing rotor to rotate smoothly.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: based on the fact that no issues could be identified and device positively passed all functional safety tests, a malfunction of the pump can be ruled out as cause of the reported event.Thus, the most likely root causes of the reported issue was an unintended user error (occlusion too hard or wrong tubing size or foreign object in the pump raceway).
|
|
Event Description
|
See initial report.
|
|