• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2021
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could not reproduce the reported issue during extensive testing of the device.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 double head pump used as sucker pump gave an error message.Reportedly it happened several times during procedure.There was no report of patient injury.
 
Manufacturer Narrative
H.10: the pump serial read out analysis revealed that on the date of the event three errors occurred.Possible root causes are: occlusion set too hard by the user.Wrong tubing size.Rotor blocked by foreign object.Bearing defective preventing rotor to rotate smoothly.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: based on the fact that no issues could be identified and device positively passed all functional safety tests, a malfunction of the pump can be ruled out as cause of the reported event.Thus, the most likely root causes of the reported issue was an unintended user error (occlusion too hard or wrong tubing size or foreign object in the pump raceway).
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
MDR Report Key12216362
MDR Text Key263283109
Report Number9611109-2021-00392
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 07/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Date Manufacturer Received09/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-