Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.Corrected data d1, d2, d4.The hakim valve was returned for evaluation: dhr - lot cpdch0, conformed to the specifications when released to stock failure analysis - the valve was visually inspected; the silicone housing was cut/torn over the distal connector, and marks were noted in the distal connector.The valve failed the tests for programming, leak and pressure.The valve passed the test for occlusion, reflux.The valve was dismantled and was examined under microscope at appropriate magnification: a small white debris was found stuck between the ruby ball and the seat of the ruby ball, this debris was putting pressure on the spring stopping the cam mechanism from moving correctly and stopping the ruby ball from being seated correctly.Root cause: - the root cause for the programing issue reported by the customer, is due to a debris interfering with the valve mechanism.- the root cause for the pressure issue noted during the investigation is due to a debris interfering with the valve mechanism.- the root cause for the cut/tear in the silicone housing and the marks in the connector noted during the investigation, is probably due to wrong handling, as noted in the ifu: silicone has a low cut/tear resistance.Do not use sharp instruments when handling the silicone valve or catheter, use shod forceps.Cuts or abrasions from sharp instruments may rupture or tear the silicone components.
|