| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577)
|
| Date of Event |
06/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medical products: vngd ti fem cr 62.5mm rt catalog # cp113616 lot # 176190, vngd ant stab brg 13x71 catalog # 189063 lot # 919050, series a pat std 31 3 peg catalog # 184764 lot # 950430.Customer has indicated that the product will not be returned because it remains implanted.Multiple mdr reports were filled for this event: 0001825034-2021-02136, 0001825034-2021-02139, 0001825034-2021-02140.According to tabers medical dictionary ¿ manipulation of a joint is a mobilization technique, sometimes involving a rapid thrust or stretching of a joint, with or without anesthesia.With anesthesia is (mua).Per clinical orthropaedics and related research (how to treat the stiff total knee arthroplasty? a systematic review), mua is used to treat and resolve arthrofibrosis (scar tissue).This procedure is performed to increase articular motion and reduce chronic pain from arthrofibrosis.The indication for manipulation under anesthesia is related to limited range of motion and stiffness due to adhesions/scar tissue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a knee arthroplasty.Subsequently, the patient had a manipulation under anesthesia due to pain, stiffness and arthrofibrosis.No revision procedure has been reported to date.Attempt for further information has been made, but no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being filled to relay additional information, which was unknown at the time of the initial medwatch.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|