• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PATIENT PACK; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PATIENT PACK; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1475
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.
 
Event or Problem Description
It was reported that the patient pack had a missing clip.The patient pack was removed from service.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for device evaluation.Product event summary: the patient pack was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned patient pack revealed a missing male buckle clip from the controller pocket.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to handling of the device and/or design issues.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PATIENT PACK
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key12217451
Report Number3007042319-2021-05445
Device Sequence Number19202489
Product Code DSQ
UDI-Device Identifier00888707000178
UDI-Public00888707000178
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2021
Device Model Number1475
Device Catalogue Number1475
Device Lot Number1699281
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2021
Initial Date Received by Manufacturer 07/01/2021
Supplement Date Received by Manufacturer07/30/2021
Initial Report FDA Received Date07/23/2021
Supplement Report FDA Received Date08/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD; 1103 VAD
Patient Age79 YR
Patient Weight88
-
-