| Model Number |
10033-004 |
| Medical Device Problem Code |
Environmental Particulates (2930)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer was dispatched to the customer site.The oil mist filter was reseated/adjusted to resolve the smoke/haze issue.Unit meets specifications and was returned to service.Asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue, and system risk analysis (sra).The device history record (dhr) was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the smoke/haze issue for the sterrad® nx unit was reviewed for the prior six months from open date and no significant trend was observed.Review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were replaced as a result of this issue.The assignable cause of the smoke/haze is likely due to the oil mist filter.The asp field service engineer reseated/adjusted the part and confirmed the sterrad® nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
A customer reported an event of a ¿smoke¿ or haze emitting from the sterrad® nx sterilizer.There was no report of any injuries or human reactions.The customer was instructed to power the unit off and discontinue use.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
|
| |
|
Search Alerts/Recalls
|