| Model Number |
D011 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded high out of range pace impedance measurements.Further evaluation is expected at the next follow up.This device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv)/respiratory sensor that is related to intermittent increases in impedance measurements.Engineering analysis and testing of returned products has identified that any repeated, small movements of the lead terminal ring can create wearing of the lead terminal ring and generate microscopic particles, which may accumulate and oxidize over time.This can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurements.Please refer to the incident description field for more information regarding the specific circumstances of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded high out of range pace impedance measurements.Further evaluation is expected at the next follow up.This device remains in service.No adverse patient effects were reported.
|
| |
|