• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES C0R50, 12X130 KII BALLOON BLNT TIP 6/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES C0R50, 12X130 KII BALLOON BLNT TIP 6/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number C0R50
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Name of procedure being performed: na.Event description: [name] incoming inspection po# 154p008.This is a complaint from [name].The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: may 31, 2021.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch - c0r50 lot# 1412557 (1 ea).Type of intervention: na.Patient status: na (no patient involvement).
 
Event or Problem Description
Name of procedure being performed: na.Event description: [name] incoming inspection po# (b)(4).This is a complaint from [name].The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: may 31, 2021.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch - c0r50 lot# 1412557 (1 ea).Type of intervention: na.Patient status: na (no patient involvement).
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair originated from an operator during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C0R50, 12X130 KII BALLOON BLNT TIP 6/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key12220974
Report Number2027111-2021-00582
Device Sequence Number11930308
Product Code GCJ
UDI-Device Identifier00607915116460
UDI-Public(01)00607915116460(17)240317(30)01(10)1412557
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K060629
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date03/17/2024
Device Model NumberC0R50
Device Catalogue Number101451901
Device Lot Number1412557
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/29/2021
Supplement Date Received by Manufacturer06/29/2021
Initial Report FDA Received Date07/23/2021
Supplement Report FDA Received Date09/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-