BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U228 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during an implant procedure, all leads were tested in the psa and no abnormal measurements were exhibited.After connecting the left ventricular (lv) lead, high out or range (oor) pacing impedance measurements were exhibited along with noise oversensing.The lv connection was cleaned and lead tested with psa again, however, oor measurements keep exhibiting.The lv lead was connected to another pg and all measurements were within range.A connection issue was suspected.The device was successfully replaced during the same procedure.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during an implant procedure, all leads were tested in the psa and no abnormal measurements were exhibited.After connecting the left ventricular (lv) lead, high out or range (oor) pacing impedance measurements were exhibited along with noise oversensing.The lv connection was cleaned and lead tested with psa again, however, oor measurements keep exhibiting.The lv lead was connected to another pg and all measurements were within range.A connection issue was suspected.The device was successfully replaced during the same procedure.No adverse patient effects were reported.This device was received for analysis.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during an implant procedure, all leads were tested in the psa and no abnormal measurements were exhibited.After connecting the left ventricular (lv) lead, high out or range (oor) pacing impedance measurements were exhibited along with noise oversensing.The lv connection was cleaned and lead tested with psa again, however, oor measurements keep exhibiting.The lv lead was connected to another pg and all measurements were within range.A connection issue was suspected.The device was successfully replaced during the same procedure.No adverse patient effects were reported.This device was received for analysis.
|
|
Search Alerts/Recalls
|
|
|