| Catalog Number |
1C8333 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported particulate matter (pm) was observed within an interlink system blood bag spike injection site.The pm was further described as white discoloration around the septum.This issue was observed after spiking prior to patient use of the device.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h3, h4, and h6.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph was reviewed, and it was noted that particulate matter (pm) contamination was observed in the connector.The pm was described as white discoloration around the septum of the device.Due to the nature of the returned sample, no additional tests could be performed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|