• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE 3 LVAS MINI APICAL CUFF KIT; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE 3 LVAS MINI APICAL CUFF KIT; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 10005877
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is completed.
 
Event or Problem Description
It was reported that when the mini apical cuff was opened the clear o-ring was discovered to be missing.This was discovered by a scrub nurse who removed the handle from the package, and upon placing it inside the mini apical cuff it came right back out.The box and package were inspected for the missing o-ring, however it was not found.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the report that the apical cuff holding tool was missing its clear o-ring (sealing ring attachment), out-of-box, could not be confirmed through this evaluation.It was reported that when the mini apical cuff was opened onto the field, the scrub nurse removed the handle from the package and went to place it inside the mini cuff, and the handle came right out.When the scrub nurse held it up, it was observed that the o-ring was missing from the end of the holder.The handle was not used.The patient did not experience any adverse consequences related to this event, and will remain ongoing with routine care.Review of the device history records (dhr) for the apical cuff holding tool assembly showed no deviations from manufacturing or quality assurance specifications.The relevant shop order found that the steps for confirming orientation on the sealing ring attachment were successfully performed for each holding tool within the lot.Although a specific duration of time for which the sealing ring was missing could not be conclusively determined, it should be noted that this does not appear to be a manufacturing-related issue based on the dhr.The clinical specialist reported that no photos of the tool were available.It was reported that the apical cuff holder would not be returned for evaluation.Review of the device history records for the mini apical cuff kit, lot number 7803980, showed no deviations from manufacturing or quality assurance specifications.Review of the dhrs for the apical cuff holding tool assembly associated with this kit, for apical cuff holding tool lot number also showed no deviations from manufacturing or quality assurance specifications.Shop order found that the steps for confirming orientation on the sealing ring attachment were successfully performed for each holding tool within lot number, per fab.The heartmate 3 (hm3) mini apical cuff kit instructions for use (ifu) is currently available.The overview section of this ifu provides a description of the apical cuff holder, stating that a silicone o-ring is present at the interface of the circular disk and apical cuff, for a hemostatic connection with the apical cuff.This section also cautions the user to notify the appropriate thoratec personnel if there is a change in how the device works, sounds, or feels.Adverse events associated with the use of the mini apical cuff kit are also outlined; however, it should be noted that the patient did not experience any adverse consequences related to the current event.The surgical procedures section of the hm3 mini apical cuff kit ifu explains how to prepare the mini apical cuff and the apical cuff holder for use.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE 3 LVAS MINI APICAL CUFF KIT
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
MDR Report Key12221775
Report Number2916596-2021-03653
Device Sequence Number14383481
Product Code DSQ
UDI-Device Identifier00813024013082
UDI-Public00813024013082
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P160054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/22/2024
Device Model Number10005877
Device Catalogue Number10005877
Device Lot Number7803980
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer10/04/2021
Initial Report FDA Received Date07/23/2021
Supplement Report FDA Received Date10/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-