| Model Number |
16122-01-S |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
06/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not returned.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
While malleting the arcus needle into the pedicle, the inner tocar cracked the outer sleeve of the jamshidi.The tocar became lodged within the bone which was very difficult to remove.The patient had very hard bone.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5.No patient injury reported.H6: medical device problem code:1260 component code:752 type of investigation: 3331; 4114 investigation findings:3252 investigations conclusions: 23 device evaluation: the device history record was reviewed.No irregularities were identified during manufacturing or inspection.The device was not returned for evaluation.A photograph was provided to confirm the event.The root cause is the result of the cannula hub interface yields/strips at a lower force than the force required to remove the product from bone.This is likely result related to the manufacturing process.
|
| |
|