Model Number PCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Glaucoma (1875); Hyphema (1911); Uveitis (2122)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2019 and (b)(6) 2021.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a pcb00 intraocular lens(iol) was explanted from patient's right eye due to patient having hyphema, glaucoma syndrome and uveitis.The replacement lens used is another johnson and johnson surgical vision lens of same model and diopter.There was no patient injury and no medical/ surgical interventions such as incision enlargement, vitrectomy or sutures were done.Patient was doing fine post-op.No further information available.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes section d9: returned to manufacturer on: 7/22/2021 section h3: device evaluated by manufacturer: yes device evaluation: the product returned in a plastic bag with a plastic container.Visual evaluation was performed, and the following were the findings.Lens came cut in a half which is associated to the explant wrapped in a gauze.The two (2) haptics were complete.Some particles were in the lens surface.The complaint issue reported was that the lens was explanted due to patient having hyphema, glaucoma syndrome and uveitis.The condition in which the sample returned is consistent with a product that was explanted.Production deficiency can¿t be determined.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.There are no discrepancies found during the mrr (manufacturing record review).A search revealed no additional investigation related to this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key12222333
Report Number2648035-2021-08173
Device Sequence Number4062559
Product Code HQL
UDI-Device Identifier05050474558243
UDI-Public(01)05050474558243(17)220328
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/28/2022
Device Model NumberPCB00
Device Catalogue NumberPCB0000190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/24/2021
Supplement Date Received by Manufacturer07/22/2021
Initial Report FDA Received Date07/23/2021
Supplement Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient Weight86
-
-