| Model Number |
EG-2990I |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: it was caused due to a physical damage on the x-ring.This report is being filed as part of the pentax backlog management plan.
|
| |
|
Event or Problem Description
|
|
The time of event is unknown.There was no report of patient harm.There is a leaking between lg connector and rotatable plug, caused by the damaged x-ring.
|
| |
|