Manufacturer's investigation conclusion: the reported event of a bent pin in the centrimag flow probe connector confirmed via evaluation of the returned flow probe (serial number: (b)(6)).Visual inspection of the returned flow probe revealed that one of the pins in the connector was bent.The returned flow probe was connected to a known working test centrimag system; however, the flow probe connector could not be fully seated to the connector on the centrimag console due to the bent pin.A flow signal interrupted: f2 alarm also activated due to the flow probe not being able to fully connect to the console.The bent pin was then straightened to continue with testing.After straightening the bent pin, the flow probe connected to the console without any issues and no atypical alarms were produced.The root cause of the reported event could not be conclusively determined through this analysis.The 2nd generation centrimag system operating manual (rev.11) section 10 ¿ ¿emergencies/troubleshooting¿ provides instructions for operation when there is a need to exchange the main console or motor with a backup console or motor.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.The 2nd generation centrimag system operating manual (rev.11) section 12.1 ¿ "appendix i ¿ primary console alarms and alerts" cover all alarms (auditory and visual), including flow alerts, and the appropriate actions to take if the issue does not resolve.The device history records were reviewed and the records revealed that the centrimag flow probe, serial number (b)(6), was manufactured in accordance with manufacturing and qa specifications.No further information was provided.The manufacturer is closing the file on this event.
|